MAINTENANCE
7 - 6
5. Clean the sensor cap and all the surface of the filter
thoroughly with methanol using a cotton swab.
6. If the filter is dirty, clean the filter with methanol.
7. Refer to Figure 7.1, re-assemble the cleaned sensor
module and the filter.
8. Make sure that the sensor module and filter is fully
dried before it is put back into the instrument.
9. Refer to Figure 7.1, insert the sensor module and filter
into the lamp housing. Make sure that the two pins on
the sensor module is aligned with the two jacks on the
lamp housing and the pogo pin of the lamp housing
goes through the hole on the sensor module. Re-install
the sensor cap to the Pocket PID Unit.