PROGRAMMING OF TOXIRAE
4 - 5
4.3 Entering into Programming Mode
1. Turn on the ToxiRAE unit (refer to Section 2.3 for the
power on sequence).
2. Press and hold down both
[MODE]
and
[N/-]
keys
for
three seconds to enter programming mode. Note: this
is to prevent user entering programming mode by
accident.
3. If the security level is 0 or 2, after holding down 2 keys
for 3 seconds, the unit should enter the programming
mode and the first menu item "Calib?" will be displayed.
If the user mode is “display”, the unit will show
"Display.." and return to normal display of
instantaneous readings.
4. If the security level is 1, after holding down the two
keys for three seconds, the display will show
"Code0000” with the left-most digit flashing. User
should enter a password starting from this flashing
digit.
Note 1:
Prior to factory shipment the ToxiRAE unit was
installed with "0000" as the default password.
Note 2:
For added security, "0000" is always displayed
instead the actual password at this step.