TROUBLESHOOTING
8 - 4
Problem Possible
Reasons & Solutions
Reading
abnormally
high
Reasons:
Dirty sensor module.
Dirty filter.
Excessive moisture and water
condensation.
Solutions:
Clean sensor module.
Replace filter.
Blow dry the sensor module.
“Lmp”/“Fan”
message
during
operation
Reasons:
Dirty sensor.
Weak or defective lamp.
Defective pump (Fan).
Solutions:
Clean sensor.
Clean or replace lamp.
Replace lamp housing with
pump.
Note:
ToxiRAE (PGM-30) with
firmware
V1.30 and above
requires replacement lamp
(
part #050-0040-000
). ToxiRAE (PGM-30)
with
firmware V1.27
and below
requires
replacement lamp (
part #050-0010-000
).
Reading
abnormally
low
Reasons:
Lamp dirty or not working.
Incorrect calibration.
Low sensitivity to the
specific vapor.
Solutions:
Replace lamp & filter.
Calibrate the unit.