PROGRAMMING OF TOXIRAE
4 - 19
4.5.3 Set Low Alarm Limit
1. The third sub-menu item in Table 4.3 is to set the Lo
Alarm limit.
- Display shows "Lo Alarm" and "+- 0050.0" with the
left-most digit flashing, where "0050.0“ is the
previously stored Lo Alarm limit.
2. To modify this limit, starting from the left-most digit,
use
[Y/+]
or
[N/-]
key to change the digit value and
press
[MODE]
key momentarily to advance to the next
digit. The flashing digit will move on to next digit to
its right. Repeat this process until all 4 digits of the
new Lo Alarm limit are entered. Press and hold
[MODE]
key for 1 second to exit data entry mode. If
there is any change to the existing value, the display
shows "Save?" Press
[Y/+]
key to accept the new value
and move to next sub-menu. Press
[N/-]
key to discard
the changes.
3.
To preserve the previously stored Lo Alarm limit, press
and hold
[MODE]
key for 1 second, the unit will exit
data entry mode and move to the next sub-menu option.