PROGRAMMING OF TOXIRAE
4 - 13
4. Turn off the flow of gas. Disconnect the calibration
adapter from the ToxiRAE Unit. Ensure that the sensor
cap is still firmly secured onto the PID unit.
5. Press the
[MODE]
key to exit the calibration sub-menu
and move directly to next menu item "Set Alarm
Limits" shown in Table 4.1.
Note:
When span check is performed, a time stamp
and the actual check value are stored in the non-volatile
memory. This information will also be included in the
datalogging report.