TROUBLESHOOTING
8 - 6
Problem Possible
Reasons & Solutions
"Bat"
message in
operation
Reasons:
Discharged battery.
Solutions:
Recharge battery.
Full scale
measurement
in humid
environment
Reasons:
Dirty or wet sensor.
Solutions:
Clean and dry sensor.
Replace filter.
Measurement
max out at
certain level
Reasons:
Dirty lamp / sensor module.
Weak lamp.
Solutions:
Clean lamp / sensor module.
Replace new lamp.
Replace filter.
Excessive
pump noise
or fan blade
does not turn
Reasons:
Defective fan blade.
Incorrect seating of fan.
Defective motor.
Solutions:
Replace new lamp housing
with pump.