OPERATION OF TOXIRAE
2 - 12
Alarm Signal Latching:
It is possible to setup the ToxiRAE from a PC so that when
an alarm condition occurs, the alarm signals stay on even
after the alarm condition is no longer present. This is
called "latching alarm" mode. The alternative mode is to
automatically reset the alarm signal when the alarm
condition is absent. This is the default alarm mode (see
Section 5.0 for details on how to set alarm mode).
Alarm Signal Summary:
The various alarm signals are summarized here:
Condition Alarm
Signal
Gas exceeds “Hi
Alarm” limit
3 beeps/flashes per second
Gas exceeds “Lo
Alarm” limit
2 beeps/flashes per second
Gas exceeds STEL
1 beep/flash per second
Gas exceeds TWA
1 beep/flash per second
Pump (fan) failure
3 beeps/flashes per second
plus "Fan" message on LCD
Sensor (lamp)
failure
3 beeps/flashes per second
plus "Lmp" message on LCD
Low battery
1 beep/flash per minute plus
"Bat" message on LCD
Memory full
1 beep/flash per minute plus
"Mem" message on LCD