PROGRAMMING OF TOXIRAE
4 - 18
4.5.2 Set STEL Alarm Limit
1. The second sub-menu item in Table 4.3 is to set the
STEL alarm limit.
-Display shows "STEL" and "+- 0025.0" with the left-
most digit flashing, where "0025.0" is the previously
stored STEL alarm limit.
2. To modify this limit, starting from the left-most digit,
use the
[Y/+]
or
[N/-]
key to change the digit value and
press
[MODE]
key momentarily to advance to the next
digit. The flashing digit will move on to next digit to
its right. Repeat this process until all 4 digits of the
new STEL alarm limit are entered. Press and hold the
[MODE]
key for 1 second to exit data entry mode. If
there is any change to the existing value, the display
shows "Save?" Press the
[Y/+]
key to accept the new
value and move to next sub-menu. Press the
[N/-]
key
to discard the changes.
3.
To preserve the previously stored STEL alarm limit,
press
and hold the
[MODE]
key for 1 second, the unit
will exit data entry mode and move to the next sub-
menu option.