Procedure:
•
Put a memory stick into USB port of a scale,
•
Enter submenu:
„
Others /
Start logo”
according to ch. 21 of this
manual,
•
Enter
<
Start logo>
to open content of a memory stick main folder,
•
Chose the desirable graphic file and wait until scales program
automatically returns to the
<
Start logo>
submenu. Chosen graphic
will be displayed.
<
Set Default>
option enables the user to get back to the default
„Start
logo”
.
Caution:
Maintained file formats are *.jpg, *.png, with optimal (max.) resolution of
640x480 pixels
.
21.11. Error information preview time
Submenu
<
Error information preview time>
enables user to declare
for how long information about errors is to be displayed.
Procedure:
•
Enter submenu
<
Others>
according to ch. 21 of this manual, select
parameter
<
Error information preview time>
and set the appropriate
option.
Where:
1 [s]
- Displaying an information about errors for 1 [s]
3 [s]
- Displaying an information about errors for 3 [s]
5 [s]
- Displaying an information about errors for 5 [s]
10 [s]
- Displaying an information about errors for 10 [s]
Max
- Displaying an information about errors until it has been
confirmed with
button.
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...