30
14.2. Filter
This filter is intended to suppress continuous mechanical vibrations at the
cost of stabilization time.
Procedure:
•
Enter
<
Weighing>
according to ch. 14 of the manual, select
<
Filter>
and then set an appropriate value.
Accessible settings:
None, V.Fast, Fast, Average, Slow.
Caution:
The higher filtering level the longer stabilization time.
14.3. Autozero
The autozero function has been implemented in order to assure precise
indications. This function controls and corrects
„0”
indication.
While the function is active it compares the results continuously with
constant frequency. If two sequential results differ less than the declared value
of autozero range, so the scale will be automatically zeroed and the pictograms
and
will be displayed.
If AUTOZERO is disabled zero is not corrected automatically. However,
in particular cases, this function can disrupt the measurement process e.g.
slow pouring of liquid or powder on the weighing pan. In this case, it is
advisable to disable the autozero function.
Procedure:
•
Enter
<
Weighing>
according to ch. 14 of the manual, select
<
Autozero>
and then set an appropriate value.
Accessible settings:
- Autozero off
- Autozero on
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...