![RADWAG PUE HY10 Скачать руководство пользователя страница 260](http://html2.mh-extra.com/html/radwag/pue-hy10/pue-hy10_user-manual_3388032260.webp)
260
A
- Product, 2 bytes, word
K
- Client, 2 bytes, word
OK
- Packages, 2 bytes, word
MZ
- Source warehouse, 2 bytes, word
MD
- Destination warehouse, 2 bytes, word
RC
- Formulation, 2 bytes, word
41.3. Description of variables
41.3.1. Output variables
Reading the output variables enables obtaining data on device status.
List of output variables:
Variable
Address Length [word]
Data type
Mass of platform 1
0
2
float
Tare of platform 1
4
2
float
Measuring unit of platform 1 8
1
word
Status of platform 1
10
1
word
Lo limit of platform 1
12
2
float
Mass of platform 2
16
2
float
Tare of platform 2
20
2
float
Measuring unit of platform 2 24
1
word
Status of platform 2
26
1
word
Lo limit of platform 2
28
2
float
Mass of platform 3
32
2
float
Tare of platform 3
36
2
float
Measuring unit of platform 3 40
1
word
Status of platform 3
42
1
word
Lo limit of platform 3
44
2
float
Mass of platform 4
48
2
float
Tare of platform 4
52
2
float
Measuring unit of platform 4 56
1
word
Status of platform 4
58
1
word
Lo limit of platform 4
60
2
float
Process status (Stop, Start)
64
1
word
Inputs status
66
1
word
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...