![RADWAG PUE HY10 Скачать руководство пользователя страница 270](http://html2.mh-extra.com/html/radwag/pue-hy10/pue-hy10_user-manual_3388032270.webp)
270
•
“LABEL EDITOR R02”
computer program,
•
"RAD-KEY"
computer program,
•
"PW-WIN"
computer program.
44. TECHNICAL PARAMETERS
PUE HY10
Housing
Stainless steel
IP rating acc. to PN-EN 60529
IP68 / IP69
Display
10,1” – widescreen display
1024 x 600 + capacitive touch screen
Power supply
100-240VAC 50-60Hz
Optional power supply
External 12-24VDC
Operating temperature
-10°C ÷ 40°C
OIML
III
Verification units quantity
6000
Max input signal
19,5mV
Max voltage per verification unit
3,25µV
Min voltage per verification unit
0,4µV
Min load cell impedance
50 Ω
Max load cell impedance
1200 Ω
Load cell power supply
5V
Load cell connection
4 or 6 wires + shield
Weighing platforms quantity
Standard 1, max.2
RS232x2
M12 8P hermetic connector
USBx2
M12 4P hermetic connector
USB A hermetic connector
Ethernet
10/100Mb, RJ45 hermetic connector
IN/OUT
4IN – M12 8P connector
4OUT – M12 8P connector
OPTIONAL EQUIPMENT
IN12 module
12IN/12OUT – gland
ANxxxx module
Current loop 4-20mA, 0-20mA,
Voltage loop 0-10V
Profibus module
2x M12 5P B-coded connector
RS485 module
Cable through gland
RS232x1
M12 8P hermetic connector
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...