![RADWAG PUE HY10 Скачать руководство пользователя страница 229](http://html2.mh-extra.com/html/radwag/pue-hy10/pue-hy10_user-manual_3388032229.webp)
229
37.4.10. Switch on continuous transmission in basic unit
Syntax:
C1 CR LF
Possible answers:
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
space
sign
mass
space
unit
CR
LF
37.4.11. Switch off continuous transmission in basic unit
Syntax:
C0 CR LF
Possible answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed
37.4.12. Switch on continuous transmission in current unit
Syntax:
CU1 CR LF
Possible answers:
CU1_I CR LF
- command comprehended but cannot be executed
CU1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...