37.4.13. Switch off continuous transmission in current unit
Syntax:
CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
37.4.14. Set lower threshold
Syntax:
DH_XXXXX CR LF
, where:
XXXXX
– mass format
Possible answers:
DH_OK CR LF
- command executed
ES CR LF
- command not comprehended (wrong mass format)
37.4.15. Set upper threshold
Syntax:
UH_XXXXX CR LF
, where:
XXXXX
– mass format
Possible answers:
UH_OK CR LF
- command executed
ES CR LF
- command not comprehended (wrong mass format)
37.4.16. Read lower threshold
Syntax:
ODH CR LF
Possible answers:
DH_MASA CR LF
- command executed
Frame format:
1
2
3
4-12
13
14
15
16
17
18
19
D
H
space
mass
space
unit
space
CR
LF
Mass
- 9 characters justified to the right
Unit
- 3 characters justified to the left
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...