53.5. Automatic adjustment time-triggered
The function enables carrying out automatic adjustment each day at a specific
day and time. You can declare up to 5 values (hours) per day (e.g. 03.00, 8.00,
13.00, 18.00, 23.00).
Procedure:
•
Enter
<
Weighing /
Automatic adjustment time-triggered>
submenu and select
<
Automatic adjustment – set time (h:m)>
parameter.
•
Enter
<
Date>
parameter and enter respective hour of automatic
adjustment start (in the following order:
hour
,
minute
).
•
Entered value is automatically set as the time of internal adjustment start.
•
To disable the function, uncheck
<
Active>
parameter (
- function
disabled,
- function enabled).