![RADWAG PUE HY10 Скачать руководство пользователя страница 227](http://html2.mh-extra.com/html/radwag/pue-hy10/pue-hy10_user-manual_3388032227.webp)
227
Example:
S I CR LF
– computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
- command done, mass value
in basic unit is returned immediately.
37.4.7. Send immediate results from all platforms in basic units
Syntax:
SIA CR LF
Possible answers:
SIA_I CR LF
- command comprehended but cannot be executed at the moment
MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF
- mass values are immediately returned from all
platforms in basic units
Frame format with mass from subsequent platforms as indicator reply:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
P
n
space
stability
space
sign
mass
space
unit
CR
LF
n
- weighing platform number
mass
- 9 characters justified to the right
unit
- 3 characters justified to the left
Example:
Let us assume that both platforms are connected to scale.
S I A CR LF
– computer command
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF
- command done, mass values from
both platforms are returned in basic units
37.4.8. Send the stable result in current unit
Syntax:
SU CR LF
Possible answers:
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...