![RADWAG PUE HY10 Скачать руководство пользователя страница 74](http://html2.mh-extra.com/html/radwag/pue-hy10/pue-hy10_user-manual_3388032074.webp)
74
21.4. Screen brightness
Scale user can change brightness of the screen in the range between
0%
and
100%
.
Caution:
The default screen brightness is set to 90%.
Procedure:
•
Enter submenu
<
Other>
as described in chapter 21 of this manual.
•
Select
<
Screen brightness>
parameter,
<Screen brightness>
editing window with an on screen keyboard is displayed.
•
Enter required screen brightness value in [%] and press
button
to confirm.
21.5. Energy saving mode
The user can activate energy saving mode.
Procedure:
•
Enter submenu
<
Others>
according to ch. 21 of this manual, select
<
Energy saving mode>
parameter and set appropriate option.
Where:
None
- Energy saving mode off
Screen dimming
- Changing screen brightness
Screensaver
- Activating screen blanking mode
Slideshow
- Activating slideshow mode
Caution:
Active screen saving mode is conditioned by:
•
Active <
Screen timeout> parameter (read chapter 21.8 of this
manual),
•
Home screen of the scale displayed,
•
Zero indication.
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...