40.4. Module for an additional platform – DP4
The module increases functionality of PUE HY10 indicators, with it an
additional platform can be supported. The module is intended to be mounted
inside the indicator. For a weighing indicator featuring DP4 module (384Rxxxx)
the additional platform is connected via cable gland mounted on a housing
cover.
40.4.1. DP4 module: technical specifications
Working temperature
-10°C ÷ 40°C
OIML
III
Verification units quantity
6000
Max input signal
19,5mV
Max voltage per verification unit
3,25µV
Min voltage per verification unit
0,4µV
Min impedance of load cell
50 Ω
Max impedance of load cell
1200 Ω
Load cell power supply
5V
Load cells wiring
4 or 6 wires + shield
Multirange function
YES
40.4.2. Colours of weighing platform signal wires
RADWAG
markings
Colour
Markings of connectors on
RADWAG A/D converter boards
+INPUT
brown
+5V
-INPUT
green
AGND
+OUTPUT
yellow
+IN
- OUTPUT
white
-IN
+SENSE
grey
+REF
- SENSE
pink
- REF
SHIELD
yellow-green
Refer to shield wiring rules
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...