244
•
command
– basic command. Setting of an appropriate command’s
byte is directly responsible for carrying out a task, in accordance with
below table:
command
bytes
task
Dec
0
scale zeroing
1
1
scale tarring
2
5
dosing process start
32
6
dosing process stop (emergency stop)
64
For instance:
0000 0000 0010 0000
– command carried out, dosing process start
•
complex command
- Setting of an appropriate command’s byte
is directly responsible for carrying out a task, in accordance with
below table:
command
bytes
task
Dec
0
setting tare value
1
1
setting
LO
threshold value
2
2
setting outputs status
4
3
setting
MIN threshold
value
8
4
setting
MAX threshold
value
16
For instance:
0000 0000 0000 0010
– command carried out, LO threshold set to value
given in LO parameter (address 10); See Table “Input variables”.
Notice:
A complex command requires setting an appropriate address parameter
from 6 to 28. See table “Input variables”.
•
Platform
– complex command parameter: number of a weighing platform.
•
tare
– complex command parameter – tare value (in calibration /
adjustment unit).
Содержание PUE HY10
Страница 1: ...USER MANUAL ITKU 88 03 08 17 EN PUE HY10 WEIGHING INDICATOR www radwag com...
Страница 2: ...2 AUGUST 2017...
Страница 13: ...13 5 WEIGHING INDICATOR STRUCTURE 5 1 External view 5 2 Main dimensions PUE HY 10 indicator dimensions...
Страница 248: ...248 Fig 21 12IN 12OUT 40 1 2 12IN OUT schematic diagrams 12 inputs 12 outputs...
Страница 268: ...268 Indicator printer CITIZEN EPSON cable Indicator ZEBRA printer cable Indicator IN OUT cable...
Страница 271: ...271 45 ERROR MESSAGES...
Страница 305: ...305...