303
Caution:
With enabled <
Automatic adjustment> parameter it is important to
periodically carry out manual internal adjustment.
53.4. Internal adjustment
Internal adjustment is carried out by means of internal adjustment weight.
Procedure:
•
Enter
<
Weighing /
Internal adjustment>
submenu,
the following
message is displayed:
•
Upon completed adjustment procedure the following message is
displayed:
•
Press
button to confirm.
Caution:
The adjustment has to be carried out when the weighing pan is unloaded and
working conditions are stable (no drafts and vibrations).