- 89 -
Mode name
– parameter, working mode name, inserted in between inverted comas. The name takes
form given on a particular balance display, it is provided in a currently selected language.
n
– parameter, decimal value determining working mode number.
n →
1 – Weighing
2 – Parts Counting
3 – Deviations
4 – Dosing
5 – Formulas
6 – Animal Weighing
8 – Solids Density
9 – Liquids Density
10 – Peak Hold
11 – Totalizing
12 – Checkweighing
13 – Statistics
CAUTION! Working modes numbering is identical for each kind of balance. The numbers are assigned to
working modes names. Some balances give only the number as a response.
An example 1:
Command:
OMI CR LF
– return accessible working modes
Response:
OMI CR LF
2_” Parts counting” CR LF
4_” Dosing” CR LF
12_”Checkweighing” CR LF
– accessible working modes are given in
return:
mode name
OK CR LF
– command carried out
An example 2:
Command:
OMI CR LF
– return accessible working modes
Response:
OMI CR LF
2 CR LF
4 CR LF
12 CR LF
– accessible working modes are given in
return:
mode number
OK CR LF
– command carried out
Set working mode
Command overview: Command sets particular working mode.
Format: OMS_n CR LF
Response options:
OMS_OK CR LF
– command carried out
OMS_E CR LF
– error in-course of command execution, no parameter or incorrect
format
OMS_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining working mode number. To see detailed description go to
OMI command
An example:
Command:
OMS_13 CR LF
– set Statistics mode
Response:
OMS_OK CR LF
– Statistics mode set
Give current working mode
Command overview: Command returns currently set working mode.
Format: OMG CR LF
Response options:
OMG_n_OK CR LF
– command carried out, response: current working mode
OMG_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining working mode number. To see detailed description go to
OMI command.
An example:
Command:
OMG CR LF
– return current working mode
Response:
OMG_13_OK CR LF
– balance operates in Statistics working mode
Give accessible units
Command overview: Command returns units available for a particular device and for a current