- 93 -
Set last digit
Format: LDS_n CR LF
Response options:
LDS_OK CR LF
– command carried out
LDS_E CR LF
– error in-course of command execution, no parameter or incorrect format
LDS_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining last digit settings
n →
1 – always
2 – never
3 – when stable
CAUTION!
The numbering is assigned to a particular last digit option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, last digit settings
are assigned to the working mode.
An example:
Command:
LDS_1 CR LF
– set last digit option to value ‘always’
Response:
LDS_OK CR LF
– ‘always’ value set
Cooperation with PUE 7.1, PUE 10 terminal
Format: NT CR LF
Response options:
ES CR LF
- command not understood (format error)
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2
3
4
5
6
7
8
9
-18
19
20
-22
23
24
-32
33
34
-36
37
38
39
40
N
T
s
pac
e
s
tabi
lit
y
m
ar
k
er
z
er
o m
ar
k
er
range
m
ar
k
er
di
gi
t m
ar
k
er
s
pac
e
m
a
s
s
s
pac
e
m
as
s
uni
t
s
pac
e
ta
re
s
pac
e
tar
e uni
t
s
pac
e
H
idden
di
gi
ts
qu
a
n
ti
ty
CR
LF
NT
- command
Stability marker
- [space] if stable, [?] if unstable
Zero marker
- [space] for any value but zero, [Z] for zero value
Range marker
- marker informing on range within which mass value is comprised:
[space] I range balance, [2] II balance, [3] III balance
Digit marker
[zero] no digit, [1] one digit marker, [2] two digits marker, [3] three
digits marker, [4] four digits marker, [5] five digits marker
Mass
10 characters for net mass given in adjustment unit (with dot and a
floating-point “-“ sign )
Mass unit
3 characters with left justification
Tare
9 characters with dot - right justification (when floating-point tare gets
turned off automatically then zero value is sent)
Tare unit
3 characters with left justification
Hidden digits quantity
Number specifying quantity of hidden digits: [space] when no digits
are hidden, [1] when 1 digit is hidden
An example:
NT CR LF
- command sent from a computer
N T _ ? _ _ 0 _ _ _ _ _ - 5 . 1 1 3 _ g _ _ _ _ _ _ _ 0 . 0 0 0 _ g _ _ _0 CR LF
– command carried
out, response: mass value and remaining data
Where: _ - space