- 84 -
Commands Overview
Zero balance
Format:
Z CR LF
Response options:
Z_A CR LF
- command understood and in progress
Z_D CR LF
- command carried out
Z_A CR LF
- command understood and in progress
Z_^ CR LF
- command understood but zeroing range is exceeded
Z_A CR LF
- command understood and in progress
Z_E CR LF
- time limit exceeded while waiting for a stable measurement result
Z_I CR LF
- command understood but not accessible at this moment
Tare balance
Format:
T CR LF
Response options:
T_A CR LF
- command understood and in progress
T_D CR LF
- command carried out
T_A CR LF
- command understood and in progress
T_v CR LF
- command understood but taring range exceeded
T_A CR LF
- command understood and in progress
T_E CR LF
- time limit exceeded while waiting for stable measurement result
T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format:
OT CR LF
Response:
OT_TARE CR LF
- command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
O
T
Space
tare
space
unit
space
CR
LF
Tare
- 9 characters with right justification
Unit
- 3 characters with left justification
CAUTION!
Tare value is always given in calibration unit.
Set tare
Format:
UT_TARE CR LF
, where
TARE
– tare value
Response options:
UT_OK CR LF
- command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
CAUTION!
Use dot in tare format as decimal point.
Содержание AS R2
Страница 28: ...28 FOOTER group of parameters enabling to declare data that is to be printed on a footer printout...
Страница 35: ...35...
Страница 70: ...70 An exemplary report adjustable volume pipette 3 volumes tested...
Страница 95: ...95...
Страница 96: ...96...