- 27 -
7.4. ADJUSTMENT REPORT PRINTOUT
At the end of each adjustment process or adjustment test, an adjustment report is generated
automatically and sent to a communication port COM 1. The content of the report is declared in menu
P5.1 CAL REPORT.
The instruction on how to declare settings for this option is described in section about printouts.
The report can be printed out via a printer connected to the balance or it can be sent to the computer
and saved as a file for archiving purposes.
8. DETERMINING PRINTOUT CONTENT
8.1. ADJUSTMENT REPORT
P5.1 CAL. REPORT,
is a group of parameters enabling user to declare data that is to be printed on
an adjustment printout.
Variable
Overview
PROJECT
Option enables naming the project (name associated with a particular type of
weighing). The name may consist of 16 characters maximum.
CALIB TYPE
Option enables printing out the type of the adjustment being carried out.
USER
Option enables printing out the name of a logged-in user.
PROJECT
Option enables printing out the name of the project (see parameter Project).
DATE
Option enables printing out the date of the carried out adjustment.
TIME
Option enables printing out the time of the carried out adjustment.
BALANCE ID
Option enables printing out the balance ID number.
CAL. DIFFER
Option enables printing out the difference between mass of an adjustment
weight measured during the last adjustment and the current measured mass
of this weight.
DASHES
Option enables printing out dashes that separate the date of a printout from
a signature.
SIGNATURE
Option enables providing an area for the signature of a user performing the
adjustment.
For the parameters described above, one of these values must
be selected:
NO -
do not print
YES -
An exemplary report:
8.2. HEADER, FOOTER, GLP PRINTOUTS
HEADER
group of parameters enabling to declare data that is to be printed on
a header printout.
GLP PRINTOUT
group of parameters enabling to declare data that is to be printed on
a measurement result printout
Содержание AS R2
Страница 28: ...28 FOOTER group of parameters enabling to declare data that is to be printed on a footer printout...
Страница 35: ...35...
Страница 70: ...70 An exemplary report adjustable volume pipette 3 volumes tested...
Страница 95: ...95...
Страница 96: ...96...