- 17 -
Press to select working mode.
F5 button of the computer keyboard.
Press to change measuring units.
PRINT/ENTER
button
Press to send measurement to a printer or a computer (PRINT).
Press to confirm selected parameter value or function (ENTER).
Press to
Zero
the balance
Press to
Tare
the balance
Press to start adjustment / calibration process immediately.
F6 button of the computer keyboard.
Press to enter the main menu of a balance.
F7 button of the computer keyboard.
Press to operate balance menu or change parameter value.
5. START-UP
When plugged to mains, the balance displays program name and number, next it proceeds to the
weighing mode.
5.1. TEMPERATURE STABILIZATION PERIOD
Before start of measuring processes, it is necessary to wait until the balance reaches thermal
stabilisation.
For balances that were stored in much lower temperatures before plugging to mains (e.g. during
winter period), thermal stabilisation period shall take at least 4 hours for PS balances, and 8 hours for
AS and APP balances. During the thermal stabilization, the indications on the display panel can
change.
It is recommended that ambient temperature changes at place of use were insignificant (slow to
change).
5.2. AMBIENT CONDITIONS STATE INDICATION
The function is intended to inform on unstable ambient conditions for a balance, it is enabled only for
AS R series balances.
The function controls dynamic temperature changes occurring in the balance during its operation. If
the variation is greater than set limit values (temperature changes speed), then a blinking
thermometer pictogram is displayed on the screen.