- 42 -
YES
NO
•
Avoid side loading, in particular side shocks:
NO
NO
The balance requires adjusting before weighing process start or in case of drastic change of ambient
conditions at a workstation.
•
Before the start of weighing procedure, it is recommended to load the weighing pan a few times
with mass close to balance max capacity,
•
Check if unloaded balance indicates “precise zero”
and whether measurement is stable
, if not, press
/Delete
button,
•
Press
UNITS
button,
to set a measuring unit:
•
Place weighed object on the weighing pan and read result only on measurement result
stabilization,
•
Mass indication of a load placed on the weighing pan can be tared multiple number of times by
pressing
→
T
←
/Insert
(pay CAUTION not to exceed maximal capacity of a balance by applying
multiple tare function).
The balance shall stay plugged to the mains in between measurement series. It is recommended to
switch off the balance display by pressing
ON/OFF
button. On repeated
ON/OFF
button pressing, the
balance is ready for operation and carrying out the following measurements.
11.2.2. Balance Zeroing
Zeroing is a function allowing to zero mass indication. In order to zero mass indication, press
button. Mass indication of zero value shall be displayed together with precise zero
and stability
markers.
Zeroing process is an equivalent for determining new zero point, recognized by the balance as
precise zero. Zeroing is possible only for stable status of display indication.
CAUTION!
Zeroing the display indication is possible only within ±2% range of instrument’s maximum capacity. If the
zeroed value is above ±2% of the maximum capacity, then the software indicates an error message, Err2.
11.2.3. Balance Taring
Taring is a function allowing to determine net weight of a measured object. In order to determine net
weight of the object, place object’s container (packaging) on the weighing pan, and on stabilization of
measurement result press
key. The display indicates mass equal zero and symbols:
Net
and
. On taking off the weighed load and its packaging from the weighing pan, the display indicates
sum of total tared mass with minus sign.
The software enables assigning tare value to a database-stored product. Using this option, the
software automatically uploads data on tare value for a particular product upon its selection from the
database.
Содержание AS R2
Страница 28: ...28 FOOTER group of parameters enabling to declare data that is to be printed on a footer printout...
Страница 35: ...35...
Страница 70: ...70 An exemplary report adjustable volume pipette 3 volumes tested...