2461Type C User Manual
Racal Instruments
1998
Programming The Type C Function Card 4-55
Interrogative form:
[:SENSe:]AVERage? ==> <ON|OFF>
Returns the state of the AVERage function.
************************************************************
[SENSe]:TINTerval:DELay:TIME <MINinum | MAXimum |
value>
Description:
Sets the time delay interval used for time interval and totalize
measurements. The selection of a delay prevents the
measurement interval from being stopped prematurely by
spurious pulses.
If <value> is specified for the time delay interval, then this will
be rounded up or down to the nearest 16 microseconds.
Parameter:
MIN = 192E-6s
MAX = 1.04856s
DEF = MIN
Initialization
*RST MIN
Example:
″
SENSe:TINTerval:DELay:TIME MAX
″
Selects maximum time delay interval.
Interrogative form:
[SENSe]:TINTerval:DELay:TIME? ==> <time
delay
interval>
Returns the time delay selected.
********************************************************************
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com