A-21
Time Interval Delay
Hold off Time
0.2 to 1.048576s
Step Size
16
µ
s
Accuracy
±
50
µ
s
±
0.1% reading.
Note:
The delay is also available on Input 2 when totalizing.
Read Rate
Frequency Measurements:
43 readings per second max. under the following conditions:
Resolution
≤
6 digits, continuous trigger, auto-trigger OFF (both channels), input freq.
≥
10kHz.
Time Interval Measurements:
43 readings per second max. under the following conditions:
Resolution
≤
6 digits, continuous trigger, auto-trigger OFF (both channels), input freq.
≥
10kHz., TI
≤
1ms
Averaging
This user selectable function can give improved resolution.
Number of Samples:
100