User Manual 2461 Type C
A-22
Availability:
Frequency 1
Frequency 2
Period 1
Period 2
Ratio 1/2
Ratio 2/1
Time Interval 1 to 2
Time Interval 2 to 1
Rise Time
Fall Time
Positive Pulse Width
Negative Pulse Width
Phase 1 relative 2
Phase 2 relative 1
Resolution Improvement:
The resolution obtained with these functions will be 1 digit more
than without averaging selected. The maximum resolution
available remains at 10 digits.
Minimum LSD on Time Interval:
0.1ns
Math Function
The following algorithm can be invoked:
(
)
scale
offset
result
−
Permitted ranges of values for offset and scale:
0, 10 x 10
-9
to 10 x 10
9
,
-1 x 10
-9
to -10 x 10
9
Permitted number of significant figures:
9
Display Range:
9.99999999X10
±
9
to -9.9999999X10
±
9
Note:
(1) Any number exceeding the allowed range for Offset or Scale will result in an error
being reported.
(2) Any number exceeding the display range will result in an error being reported.
(3) The Math Function is not available on CHECK, PHASE , PULSE WIDTH,
RISE/FALLTIME, READ MID, READ MAX, READ MIN.