Performance Verification and Adjustments 6-9
Table 6-5, Time Interval 1-2 Test Performance Limits
Input 1 Slope
Command Syntax
Input 2 Slope
Command Syntax
Results Tolerance
INP1:COMP:SLOP POS
INP2:COMP:SLOP POS
0ns ±
2ns
INP1:COMP:SLOP NEG
INP2:COMP:SLOP POS
50ns ±
2ns
INP1:COMP:SLOP NEG
INP2:COMP:SLOP NEG
0ns ±
2ns
INP1:COMP:SLOP POS
INP2:COMP:SLOP NEG
50ns ±
2ns
Command Syntax:
MEAS:TINT?
takes a Time Interval measurement
Disconnect the test equipment.