2461Type C User Manual
Programming The Type C Function Card 4-56
Racal Instruments
1998
[SENSe]:TINTerval:DELay[:STATe] <ON | OFF>
Description:
Enables or disables the time interval delay facility. The selection
of a delay prevents the measurement interval from being
stopped prematurely by spurious pulses.
Initialization:
*RST OFF
Example:
″
SENSe:TINTerval:DELay ON
″
Selects time interval delay on.
Interrogative form:
[:SENSe]: TINTerval:DELay?: ==> <ON | OFF>
Returns the state of the time interval delay
selection.
****************************************************************
[SENSe]:TOTalize:GATE[:STATe] <ON | OFF>
Description:
Enables Automatic Totalize (Totalize 1 by 2 or Totalize 2 by 1)
or Manual Totalize. When ON, Automatic Totalize is selected.
When OFF, Manual totalize is selected.
This function is only available if the currently selected function is
TOTalize. Sending this command while another measurement
function is selected generates the error –240 (“Function not
allowed”).
Initialization:
*RST ON
Example:
“CONF2:TOT”
″
SENSe:TOTalize:GATE
ON
″
Selects the Totalize 2 by 1 measurement
function..