2461Type C User Manual
Racal Instruments
1998
Programming The Type C Function Card 4-53
″
TINT
″
″
TOT
″
″
PHAS
″
″
CHEC
″
Note:
If the specified channel is different from the channel
previously configured, the 2461 TYPE C returns an error
(
″
Channel not configured
″
).
*************************************************************
[SENSe][:PERiod]:APERture MINimum | MAXimum | DEFault |
<value>
[SENSe][:FREQuency]:APERture MINimum | MAXimum |
DEFault | <value>
[SENSe][:FREQuency:RATio]:APERture MINimum | MAXimum |
DEFault | <value>
[SENSe][:TINTerval]:APERture MINimum | MAXimum | DEFault
| <value>
[SENSe][:RTIMe]:APERture MINimum | MAXimum | DEFault |
<value>
[SENSe][:RISE:TIME]:APERture MINimum | MAXimum |
DEFault | <value>
[SENSe][:FTIMe]:APERture MINimum | MAXimum | DEFault |
<value>
[SENSe][:FALL:TIME]:APERture MINimum | MAXimum |
DEFault | <value>
[SENSe][:PWIDth]:APERture MINimum | MAXimum | DEFault |
<value>
[SENSe][:NWIDth]:APERture MINimum | MAXimum | DEFault |
<value>
[SENSe][:PHASe]:APERture MINimum | MAXimum | DEFault |
<value>
[SENSe][:TOTalize]:APERture MINimum | MAXimum | DEFault
| <value>
[SENSe][:CHECk]:APERture MINimum | MAXimum | DEFault |
<value>
Description:
These commands all do the same thing. They set the 2461
TYPE C measurement aperture <time>. The units are seconds
and range from 1E-6 to 10. If the user specifies a value, then
this will be rounded up or down to the nearest of 1
µ
s, 10
µ
s,
100
µ
s, 1ms, 10ms, 100ms, 1s or 10s.
Note that the aperture for the phase and totalize function is
determined by the input signal alone.
The resolution setting may be changed indirectly by the
CONFigure:ARRay and MEASure:ARRay commands and by
the SENSe:RESolution command.