2461Type C User Manual
Racal Instruments
1998
Programming The Type C Function Card 4-11
PARAMETER:
The allowed range for <value1> and <value2> is:
1E-9 < value < 1E10
0
-1E10 < value < -1E-9
INITIALIZATION:
*RST value1 = 0 (offset)
value2
=
1
(scale)
EXAMPLE:
″
CALCulate:MATH:EXPRession 1000,MEASure
″
Specifies an offset of 1000 and instructs the 2461 TYPE C to
use the previous measurement as a scale factor.
INTERROGATIVE FORM:
CALCulate:MATH:EXPRession? ==> <offset value, scale
value>
Returns the offset and scale values currently selected.
*******************************************************
CALCulate:STATe <OFF | ON>
DESCRIPTION:
This command enables or disables the math processing that
is specified by the CALCulate:MATH:EXPRession command.
INITIALIZATION:
*RST OFF
EXAMPLE:
"CALCulate:STATe ON"
Scale
Offset
sult
−
Re
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com