Programming The Type C Function Card 4-47
PARAMETERS:
DEFault = POS
INITIALIZATION:
*RST POS
INTERROGATIVE FORM
INPut[1 | 2]:COMParator:SLOPe? ==> <POSitive |
NEGative>
Returns the current comparator slope setting for the specified
channel.
If the function RISE:TIME is active, the query returns POS.
If the function FALL:TIME is active, the query returns NEG.
If the function PWIDth is active, the query returns POS if
channel 1 is specified, or NEG if channel 2 is specified.
If the function NWIDth is active, the query returns POS if
channel 2 is specified, or NEG if channel 1 is specified.
********************************************************
INPut[1 | 2]:COMParator:SETup:AUTO <ON | OFF | ONC
DEFault
>
DESCRIPTION:
Enables, either continuously or singly, or disables the auto
trigger.
If OFF, ONCE or DEFault are specified while the functions
PHASe, RISE:TIME, FALL:TIME, PWIDth and NWIDth are
active, then the error –240 is generated (“Function Error”).
If OFF, ONCE or DEFault are specified while the functions
VOLT:MAX, VOLT:MIN and VOLT:MIDD are active and the
channel specified is the same as the channel currently
configured for the measurement function, then the error –240
is generated (“Function Error”).