User Manual 2461 Type C
A-18
Phase Range
0.1 ° to 360 °
LSD
Fin
≤
1MHz:
0.1 °
1MHz < Fin
≤
10MHz:
1 °
10MHz < Fin
≤
100MHz:
10 °
Resolution
°
×
±
±
360
1
Period
Resolution
TI
LSD
Accuracy
°
×
±
±
360
1
Period
Accuracy
TI
LSD
Note:
(1) A constant signal is required during a phase measurement.
(2) A phase measurement will nominally take 160ms at 100MHz and
4s at 1Hz.
Phase 2 Relative to 1
To make this relative phase measurement, the UCT first makes a series of
measurements in different modes:
(1)
Ratio 2/1 to ensure that 2 and 1 are at the same frequency.
(2)
Period 2 (P
2
).
(3)
Time Interval 2 to 1 (TI
2-1
).
The phase difference
Φ
2-1
is then calculated as follows:
°
×
=
Φ
−
−
360
P
TI
1
2
1
2
2