2461Type C User Manual
Performance Verification and Adjustments 6-14
Racal Instruments
1998
Functional Checks
Input 1 Filter
Check
Set the UCT to its default state. Select 50
Ω
impedance for Input
1.
Configure the UCT as follows:
*RST
returns unit to default state
INP1:IMP 50
selects 50
Ω
impedance for Input 1
Connect the UCT to the signal generator as shown in Figure 6-7.
Figure 6-7, Input 1 Filter Check Configuration
Set the generator to 50kHz, 1Vrms. Check that the UCT is
measuring 50kHz and the peak to peak signal level is:
Command Syntax Notes
MEAS1:FREQ? Confirm 50kHz
MEAS1:VOLT:MAX? V
P1
MEAS1:VOLT:MIN? V
N1
Switch in the input filter for channel 1 and again check that the
counter measures a frequency of 50kHz. Measure the peak to
peak signal and verify that the signal amplitude is reduced by
greater than 2dB.
INP1:FILT:LPAS:STAT ON
turn on LPF for channel 1
MEAS1:FREQ?
Confirm
50kHz
MEAS1:VOLT:MAX?
V
P2
MEAS1:VOLT:MIN?
V
N2
GATE
IN
1
VT1
VT2
ARM
T2
IN
STD
OUT
IN
2
OSC
UCT
IN 3
SIGNAL GENERATOR
RF OUT
INPUT 1