2461Type C User Manual
Programming The Type C Function Card 4-58
Racal Instruments
1998
This function is only available if the currently selected function is
TOTalize. Sending this command while another measurement
function is selected generates the error –240 (“Function not
allowed”).
INITIALIZATION:
*RST OFF
EXAMPLE:
″
SENSe1:TOTalize:INTermediate ON
Enables totalize on the fly measurements.
INTERROGATIVE FORM:
SENSe1:TOTalize:INTermediate? ==> <ON | OFF>
Returns the current state of the totalize on the fly
function.
*******************************************************
[SENSe[1 | 2]]:TOTalize:MANual <ON | OFF>
Description:
The Manual Totalize function enables the counting of events on
either channel 1 or 2. The counting is started with the
SENSe:TOTalize:MANual ON command, and it is stopped with
the SENSe:TOTalize:MANual OFF command. The counting of
the manual totalize function is cumulative. The count is reset by
the SENSe:TOTalize:GATE OFF command.
Note
: That the readings on-the-fly are automatically enabled.
This function is only available if the currently selected function is
TOTalize. Sending this command while another measurement
function is selected generates the error –240 (“Function not
allowed”).
Initialization:
*RST OFF
Example:
CONFigure2:TOTalize
Selects TOTalize function;
SENSe:TOTalize:GATE OFF Enables Manual
Totalize;
SENSe2:TOTalize:MANual ON Starts the counting