OPERATING INSTRUCTIONS
/ HOW TO REMOVE AND LOAD EACH RESOURCE
77
E
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
7.3.7
REMOVING THE SOLID WASTE BAG
This section explains how to empty the solid waste container of the
MosaiQ 125
. The solid
waste container is located in waste drawer. Since this waste is of a biological nature
(blood samples), a biohazard waste bag is used to line the waste container. The solid
waste drawer can be opened and closed manually when the
MosaiQ 125
is in operation.
NOTICE
Use commercially available 50 liter biohazard waste bags.
There is a waste buffer with a capacity of 10 microarrays, allowing up to 4 minutes to
replace the bag during operation.
1 - Tap on the
Resources
icon in the navigation bar.
2 - Tap on the
Liquid Waste / Solid Waste
area to check the level.
3 - Open the waste drawer.
4 - Pull out the solid waste container.
5 - Pull the rear holding clamp and release the full bag.
6 - Remove the full bag.
7 - Dispose of the full bag according to your laboratory biohazard waste procedure.
(removing of the solid waste bag)
During the emptying and loading operation, on the
Resources
screen, an icon shows that
the waste drawer door is open and the remaining quantity of microarrays in the waste
buffer container.
Wast drawer
door is open
Remaining quantity of
waste (microarrays)
(solid waste drawer status and back up bin capacity)