TECHNICAL DATA
/ ENVIRONMENTAL CONDITIONS
169
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
18.6 ENVIRONMENTAL
CONDITIONS
The following table shows the range of environmental conditions required to operate the
system safely.
Environmental Conditions
The system is made for Indoor use only.
Temperature
Operating: 16°C to 32°C (60.8°F to 89.6°F)
Storage: 5°C to 40°C (41°F to 104°F)
Transport: -10°C to 55°C (14°F to 131°F)
Humidity
Operating: 15% - 80% non-condensing
Storage: 10% - 85% non-condensing
Transport: 10% - 90% non-condensing
Pollution Degree
2
Installation Class
2
Limit Class
Class A
(For industrial use. Domestic use restricted)
Sunlight
No direct sunlight
May mislead optical sensors and affect performance
Altitude
Up to 2000 m (6562 ft) above mean sea level
Dust
No excessive dust
18.7 NOISE
This section provides information about noise generated by the
MosaiQ 125
.
47 dBA is the maximum noise level measured at 1 m (39.4 in) around the device
59 dBA is the maximum noise level in the user’s working area at 25 cm (9.8 in) from the
device.
18.8 PACKAGING
This section provides information about how the
MosaiQ 125
is packaged as well as the
dimensions and weight of each crate.
Number of crates: 2
Crate1
Contents
MosaiQ 125
Dimensions (WxDxH)
180 cm x 105 cm x 200 cm (70.87 in x 41.34 in x 78.75 in)
Weight
500 kg (1102.4 lb)
Crate 2:
Contents
Covers
Dimensions (WxDxH)
180 cm x 80 cm x 120 cm (70.87 in x 31.5 in x 47.25 in)
Weight
200 kg (441 lb)
18.9 MOSAIQ 125
MEMORY
This section provides information about the memory capacity of the
MosaiQ 125
. Memory
enables the
MosaiQ 125
to store 30 000 test results information.