MOSAIQ 125 PARTS DESCRIPTION AND LOCATION
/ PROCESS MODULES
42
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
5.11.1 REAGENT
CARRIAGE
The reagent carriage in the process modules compartment houses up to eight reagent
bottles. The symbols of the MosaiQ system reagents indicate the position where each
reagent bottle shall be inserted.
(reagent carriage localization - reagent door opened)
Function:
To house MosaiQ detection and Enhancement reagent bottles.
Capacity:
The reagent carriage can carry up to eight reagent bottles (2 bottles of each kind of
reagent). The MosaiQ system reagent are:
භ
MosaiQ Detection Reagent 1
භ
MosaiQ Detection Reagent 2
භ
MosaiQ Enhancement Reagent 1
භ
MosaiQ Enhancement Reagent 2
Technology:
An RFID reader is used to read and write on the RFID tags present on each reagent bottle.
One yellow LED is located on the top right side of each slot on the reagent carriage.
LEDs indicate the actual reagent bottle status.
(LEDs indicating the bottles status)
MOSAIQ DETECTION AND ENHANCEMENT REAGENT BOTTLES LED STATUS
Off
Position available
On
Bottle in use
Blinking slow
Identifies loading position for required reagent when necessary
Blinking fast
Error
To remove and load the MosaiQ detection and MosaiQ enhancement reagent
bottles, please refer to
chapter 7.3.3
and
chapter 7.3.4
of this manual.
To clean the reagent carriage (weekly task), please refer to
chapter 12.6.1
of this
manual.