MAINTENANCE
/ CLEANING SOLUTIONS TO BE USED
132
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
12.3.1
BEFORE PERFORMING MAINTENANCE TASKS
Check that the minimum volumes for the resources for the
MosaiQ 125
have been met,
before performing the maintenance tasks.
Reagents and system liquid,
Available capacity in the waste canister.
For more information to check resources levels please refer to
chapter 7.2.2
.
12.3.2
CLEANING AND DISINFECTION STEPS
This paragraph contains the information and instructions for the proper cleaning and
disinfection of the
MosaiQ 125
to be carried out by the user. Personal Protective Equipment
(PPE) must be worn in accordance with local regulations and guidance when performing
the
MosaiQ 125
maintenance.
Comply with the following aspects during cleaning and disinfection operations. Failure to
do so can result in breakdowns or damage to the
MosaiQ 125
and may expose a risk to the
health of individuals.:
Do not use the cleaning solution near circuit boards, light barrier or barcode reader.
Do not spray the cleaning solution inside the
MosaiQ 125
.
Always spray the cleaning solution onto a soft lint-free wipe to clean internal parts
of the
MosaiQ 125
.
You may spray the cleaning solution on the external surfaces where it is safe to do so
(away from all ports, outlets and connections). Another option is to spray cleaning
solution onto a lint-free wipe, which can then be used to clean the
MosaiQ 125
.
Do not autoclave containers and components for the system liquid or waste.
Make sure that before cleaning the
MosaiQ 125
, all samples, reagents and
magazines have been removed.
The users with the required access rights should perform each maintenance task in
accordance with the recommended frequency as described below:
12.4 CLEANING SOLUTIONS TO BE USED
The
MosaiQ 125
is designed such that the majority of cleaning activities can be performed
following local guidelines and common laboratory cleaning agents.
We recommend using either 70% Ethanol (EtOH) or Isopropyl alcohol (IPA) for cleaning of
external surfaces and 0.5 M Sodium Hydroxide (NaOH) for pipettor cleaning.
All cleaning solutions for internal use should instead be sprayed onto a lint free wipe
outside the
MosaiQ 125
first and then target surfaces wiped gently.
Use only the following products for cleaning and disinfection purposes:
70% alcohol (i.e. Isopropanol),
2% Decon® / Virkon®,
5% Bleach,
0.5 M NaOH (Sodium Hydroxide),
DI Water (Deionized Water),
Cleaning wipes / Lint free multi-surface wipes.
WARNING
Do not use bleach for cleaning of all internal
MosaiQ 125
surfaces.
The impact of bleach vapor may have an impact on functional
testing.
WARNING
Never mix cleaning products.