MAINTENANCE
/ WEEKLY MAINTENANCE
141
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
12.6.3
CLEANING OF THE REAGENT DRAWER BLACK INSERT AND
REAGENT BOTTLE DOCKING STATION
To clean the inside of the reagent drawer:
1 - Open the reagent drawer.
2 - Remove any reagent bottle from the drawer.
3 - Wipe the surface of the black insert of the reagent drawer with a lint free wipe and
70% alcohol.
4 - If any salt crystal is present, use warm water to remove it.
5 - Close the reagent drawer.
(inside of the reagent drawer)
More information about the reagent drawer, please refer to
chapter 5.8
of this manual.
To remove the reagent bottles from the reagent drawer, please refer to
chapter 7.3.5
of
this manual.
12.6.4
CLEANING OF THE EXTERNAL SURFACES OF THE MOSAIQ 125
The external surfaces (covers and drawers) can be cleaned when the
MosaiQ 125
is not
operating.
External
MosaiQ 125
covers spray 70% alcohol on a lint free wipes then clean the
MosaiQ 125
surface, however care should be taken when cleaning the
MosaiQ 125
touch screen and
PC.
1 - Close all drawers.
2 - Wipe the surface of the external covers and drawers with a lint free wipe and 70%
alcohol.
12.6.5
CLEANING OF THE REAGENT CARRIAGE
The reagent carriage is located in the process module compartment and supports the
detection and enhancement reagent bottles. It should be cleaned weekly.
WARNING
Disconnect the
Mosaiq 125
power cord from the mains power
supply before carrying out this cleaning/disinfection operation..
To clean the reagent carriage:
1 - Be sure that the
MosaiQ 125
is not connected to the mains power supply.
2 - Open the reagent door.
3 - Remove all detection and enhancement reagent bottles from the reagent carriage.
4 - Wipe all accessible surfaces of the reagent carriage with a lint free wipe and 70%
alcohol.
More information about the reagent carriage please refer to
chapter 5.9.1
of this manual.
To remove detection and enhancement reagent bottles form the reagent carriage please
refer to
chapter 7.3.3
of this manual.
12.6.6
CLEANING OF THE EXTERNAL COVERS OF THE INCUBATION
RING
Clean the incubation ring covers weekly.
To clean the incubation ring covers:
1 - Be sure that the
MosaiQ 125
is not connected to the mains power supply.
2 - Open the reagent door.
3 - Wipe the surface of the incubation ring covers with a lint free wipe and 70% alcohol.
More information about the incubator please refer to
chapter 5.9.3
of this manual.
12.6.7
CLEANING OF THE REAGENT DOOR
The reagent door is controlled by software and cannot be opened when the
MosaiQ 125
is operating, but it can be opened when the
MosaiQ 125
is in
Idle
mode. For safety reasons
disconnect the
Mosaiq 125
from the mains power supply.
WARNING
Disconnect the
Mosaiq 125
power cord from the mains power
supply before carrying out this cleaning/disinfection operation..
To clean the reagent door:
1 - Be sure that the
MosaiQ 125
is not connected to the mains power supply.
2 - Open the reagent door until it reaches its locked open position.
3 - Wipe the internal surfaces of the reagent door with a lint free wipe and 70% alcohol.
4 - Close the reagent door and wipe the external surface of the reagent door with a lint
free wipe and 70% alcohol.