MAINTENANCE
/ DAILY MAINTENANCE TASKS
135
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
12.5.2 INITIALIZE
SYSTEM
This maintenance task is required prior to any sample processing. It takes up to 5 minutes
to initialize all components. Initialization of the
MosaiQ 125
happens automatically after
switching it on. This task is mandatory.
On the touch screen:
1 - Tap on
Maintenance
on the navigation bar.
2 - Tap on
Init System
.
3 - Tap on
Execute Task
on the right side of the screen to start the maintenance.
NOTICE
The initialization of the
MosaiQ 125
is required every 8 hours.
Empty the liquid waste containers.
12.5.3 PRIME
DISPENSERS
This maintenance task primes all dispensers on the reagent carriage: Each bottle is
primed 6 times. The initialization starts automatically after switching on the
MosaiQ 125
.
This task is mandatory.
•
MosaiQ Detection Reagent 1
•
MosaiQ Detection Reagent 2
• MosaiQ Enhancement Reagent 1
• MosaiQ Enhancement Reagent 2
Priming must be repeated if a new reagent bottle is installed. Prime each (Detection,
Enhancement) reagent after inserting a new bottle.
On the touch screen:
1 - Tap on
Maintenance
on the navigation bar.
2 - Tap on
Prime Dispensers
.
3 - Tap on
Execute Task
on the right side of the screen to start the maintenance.
12.5.4
INITIAL WASHER PRIME
This maintenance task primes all washer lines and refills the intermediate container
with diluted wash buffer. It needs to be executed after each initialization. Please ensure
MosaiQ wash buffers and
MosaiQ sample diluent
s are plugged in.
On the touch screen:
1 - Tap on
Maintenance
on the navigation bar.
2 - Tap on
Initial Washer Prime
.
3 - Tap on
Execute Task
on the right side of the screen to start the maintenance.
12.5.5
PRIME WASH MODULE 1
This maintenance task primes the wash module 1, to remove any air bubbles. To be
performed as needed.
On the touch screen:
1 - Tap on
Maintenance
on the navigation bar.
2 - Tap on
Prime Wash Module 1
.
3 - Tap on
Execute Task
on the right side of the screen to start the maintenance.
12.5.6
PRIME WASH MODULE 2
This maintenance task primes the wash module 2, to remove any air bubbles To be
performed as needed.
On the touch screen:
1 - Tap on
Maintenance
on the navigation bar.
2 - Tap on
Prime Wash Module 2
.
3 - Tap on
Execute Task
on the right side of the screen to start the maintenance.
12.5.7
PRIME WASH BUFFERS AND SAMPLE DILUENTS
This maintenance task primes the fluid lines of
MosaiQ Sample Diluent 1
and
2
and it
generates fresh diluted
MosaiQ Wash buffer 1
and
2
in the intermediate bottles. To be
performed as needed.
Please ensure that all 4 bottles are plugged in.
On the touch screen:
1 - Tap on
Maintenance
on the navigation bar.
2 - Tap on
Prime Wash Buffers and Sample Diluents
.
3 - Tap on
Execute Task
on the right side of the screen to start the maintenance.
12.5.8
PRIME SAMPLE DILUENT 1
This maintenance task primes the
MosaiQ Sample Diluent 1
, to avoid or remove any
bubbles from the lines. To be performed as needed.
For an optimal result please perform this task three times consecutively.
On the touch screen:
1 - Tap on
Maintenance
on the navigation bar.
2 - Tap on
Prime Diluent 1
.
3 - Tap on
Execute Task
on the right side of the screen to start the maintenance.