116
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
10. MOSAIQ 125 SHUTDOWN
To shut down the
MosaiQ 125
safely, please follow this procedure.
10.1. REMOVAL OF REMAINING MATERIAL AND WASTE
1 - Remove all racks from the
MosaiQ 125
.
(removing the sample rack)
2 - Remove all samples from the racks.
(removing the sample tubes)
3 - Empty the liquid canister and solid waste container, if necessary. Discard the waste following
the procedure.
(removing the liquid and solid wast)
4 - Remove any empty magazines from the carousel.
(removing the remaining magazines)
5 - Remove any empty reagents bottles.
(removing the remaining reagent bottles)