Appendix
Messages during balancing
Insufficient improvement
The vibration values are greater than 1 mm/s and have not
sufficiently improved. Check whether the cause is actually an
imbalance.
One-plane balancing is recommended
The dynamic portion of the calculated balancing weight is larger than
the static portion. The results will not be significantly improved if you
proceed to balance in two planes.
Remove balancing weight?
You can leave the balancing weight on the rotor or remove it after
each balancing run. Answer the question with YES or NO.
RPM measurement
VIBXPERT measures the RPM before each balancing run. The green
LED lights up if the RPM sensor sends a pulse.
Please check trigger sensor!
There is no signal from the trigger / RPM sensor. Possible causes:
signal path is interrupted, poor light conditions, incorrect orientation
of the trigger optics for the reference mark, ...
Initializing
The measurement amplifier is adjusted (Autorange)
Speed fluctuates. Results can be influenced!
The RPM variation is too large or deviates too much from the
measured RPM of the preceding balancing run. Wait until the
machine has reached the balancing RPM before starting the measure-
ment.
Vibration pointer unstable
The magnitude and/or the phase of the vibration pointer is changing
too much. VIBXPERT automatically increases the selected number of
averages until the signal is stable or the maximum number of
averages (20) is reached.
Maximum number of averages reached
VIBXPERT has set the number of averages to 20. If the vibration
pointer has stabilized, the balancing runs can be continued.
Balancing planes cannot be separated (2-plane balancing)
The influences of the balancing weights in the other plane are the
same or are adjusted so that they increase at the same time. Under
these conditions, the results for the subsequent balancing runs
cannot be used. Balancing in one plane is recommended.