2-26
VIBXPERT Balancing 11.2010
Setup parameters
This section describes the measurement settings that you generally
can only edit before beginning a balance measurement. The parame-
ters are contained in the 'Setup menu':
- Machine setup (see below)
- Measurement setup (see p. 2-28)
- Transducer setup (see VIBXPERT operating manual, page 2-20)
Machine setup
A 'machine setup' contains the machine data and information on the
measurement setup. VIBXPERT uses these data to calculate a suitable
trial weight and, after each balancing run, the remaining imbalance,
centrifugal force and the attained balancing quality. To create a
machine setup, proceed as follows:
• Mark a rotor type in the selection screen (see p. 2-11).
• Press the MENU button and click on "Task manager".
• Mark the "Machine setup" field.
• Press the MENU button and click on "New" or on "Edit":
P
LANES
:
One / Two
. Number of balancing planes; cannot be changed
because the rotor type is selected in the selection screen.
C
ORRECTION
MODE
*:
Free / fixed location / fixed weight / meas. tape
;
takes into account restrictions on the machine (fixed locations, e.g.
for ventilator blades) or local circumstances (tape measure).
Depending on the selection, the following parameters can be
selected:
- Number of locations: 3-100. Usually the number of ventilator
blades.
Balancing: Machine setup
*This parameter can be changed in
each balancing run (MENU - Tools)