2-14
VIBXPERT Balancing 11.2010
Balancing: 1-plane
Note
In the following trim runs VIBXPERT checks whether the imbalance
has sufficiently reduced from one run to the next. If this is not the
case, the following message appears: ‘No Improvement’. The run
concerned must be repeated with a different balancing weight.
Start again if the vibration values worsen and the balancing
weights do not become smaller.
8. Move the joystick to the right and continue the balancing with the
next trim run.
The balancing procedure is complete as soon as the vibration values
have reached an acceptable value. When working with "Machine
Setup", the calculated balancing quality is the criterion for terminat-
ing the balancing procedure. A ‘smiley’ appears in the measurement
screen if the balancing quality has reached the tolerance range.
Undoing balancing runs
If the measurement results worsen after a trim run, you can return to
a run that was acceptable and continue balancing with different
weights from there.
• Navigate to the data screen of the trim run at which you wish to
continue the procedure.
• Press the MENU button and click on 'Undo' (see below).
• Confirm the next query with 'Yes'. All subsequent trim runs are
deleted.
• Remove the balancing weights that you attached during the
deleted trim runs.
* = Trim weight was removed
Left:
Balancing quality achieved
(balancing with machine setup)
Right:
Undoing balancing runs