Fig. 44: [0.02 NTU calibration] display
A0527
Fig. 45: [Start calibration] display
10. Insert the 0.02 NTU standard solution. Rotate the standard
solution in the measuring chamber so that the lowest value is
displayed in the upper line of the display.
11. Press EXIT to start calibration.
ð
The number counts down from 30 in the lower display.
A3422
Fig. 46: [CAL Good]
12. The device displays
[CAL Good] at the end of the process.
13. Press EXIT to confirm. The display briefly shows
[STRT] and
then changes to
[AUTO].
7.1.5 Aborting calibration
1.
You can abort calibration at any time during the calibration
process by pressing the DOWN key until
‘ABRT’ is dis‐
played. Then press EXIT.
ð
The device returns to AUTO mode.
2.
The device saves and uses all points that have been suc‐
cessfully calibrated. Note that the calibration log records any
missing calibration values or failed calibrations.