The system can detect the following fault statuses:
n
Incorrect cuvette installed
n
An error has occurred in the transducer
n
The transducer has no contact with the spring
If an error occurs, the message
[CLN] appears in the lower part of
the LCD display. An alarm is triggered and 2 mA is applied to the
current output.
The alarm relay is triggered if the alarm function in
[ALM1/ALM2] is
set to
[ERROR]
The 4 ... 20 mA output is set to 0, 2, 4 mA if the
[error level] [ERLV]
has been set to the corresponding mA level (0, 2, 4 mA).
3.4 RS-485 interface
The sensor offers basic communication functions using programs,
such as
[Hilgraeve HyperTerminal] (included in most
[Microsoft Windows software] packages) etc. [Visual Basic] or
other programs can also be used.
3.5 Scope of delivery
INFORMATION: Packaging material. Dispose of all packaging
material in an environmentally responsible way. All packaging com‐
ponents carry the corresponding recycling code .
INFORMATION: A packet of dessicant is included in the scope of
delivery of the DULCO turb C. Do not mistake this for packaging
material. Do not throw away the dessicant.
Tab. 2: The following parts are delivered as part of the standard scope of delivery of a DULCO turb C:
Designation
Quantity
DULCO turb C with terminal box and integrated flow fitting
1
Operating instructions
1
Packet of dessicant
1
Cuvette
1
Hose kit:
n
Shut-off clamp (1x)
n
Back pressure valve (1x)
n
Connection hoses with connection pieces for the flow assembly (2x).
1
USB stick (contains files for optional ranges and the operating instructions).
1
Remove the DULCO turb C from its packaging box. Check all com‐
ponents. Ensure that no visible damage has occurred during trans‐
port. Immediately contact your local distributor or the ProMinent
Service department if the components supplied fail to comply with
the order.
Functional description / Product identification
12