You need to enter the access code to access the
‘CAL’ or
‘CONFIG’ menus.
1.
When the security function is activated (“ON”), press the
RIGHT key.
A3415
Fig. 14: Screen for inputting the access code
ð
The screen in which the access code is entered appears.
2.
The first number of the access code flashes. Use the UP or
DOWN key to select the correct number and confirm it with
the EXIT key.
ð
The second number of the access code flashes.
3.
Use the UP or DOWN key to select the correct number and
confirm it with the EXIT key.
ð
The third number of the access code flashes.
4.
Use the UP or DOWN key to select the correct number and
confirm it with the EXIT key.
ð
If you have selected the valid access code, then you will
now have access to the sensor's calibration mode. In the
event that the access code is incorrect, the sensor
returns to AUTO mode.
6.5 Device configuration
6.5.1 Selecting the output
A0530
Fig. 15: Selecting the output
Entering the access code
Commissioning
33